Auto Chemistry Analyzer BK-280

 

Introduction:

It is used to quantitatively analyze the clinical chemistry of human serum, plasma, urine, cerebrospinal fluid and other samples. Do not use for other purposes.

 

Festures:
1. 49 sample positions.
2. 56 reagent positions.

3. 120 reaction cuvettes.

4. 200 tests per hour.

5. Probe with anti-collision function, liquid level dectection function.

Description
Product Specification

Technical Parameters:

ModelBK-280

Overall

Performance

Throughput200 tests/hour
Analysis MethodEnd-point, Fixed-time, Rate(Kinetic), Turbidimetry
CertificatesCE, FDA, ISO9001, ISO13485
Sample&Reagent UnitSample Positions49 sample positions
Reagent Positions56 reagent positions
Sample Volume2~70μl
Reagent Volume20~350μl
ProbeTeflon coating, with anti-collision function, liquid level detection function
Probe WashingAutomatic washing interior and exterior
Reagent CoolingRefrigerated tray with independent switch
Reaction SystemTemperature Control37±0.2℃
Cuvettes120 reusable cuvettes, optical length 6mm
MixerProbeIndependent stirring
WashingAutomatic cuvettes washing
STAT functionYES
Optical SystemLight Source12V/30W halogen lamp
SpectrophotometryPost-spectral spectrophotometry
Wavelength340, 405, 450, 480, 505, 546, 570, 600, 630, 700, 750, 800nm
Absorbance0~3.0Abs
Calibration&QCCalibration

Linear: K factor, 1-point, 2-point and multipoint linear

Non-Linear: Spline, Polygon, Logit-4P, Logit-5P

Quality ControlReal-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D), Daily QC, Monthly QC
Data ManagementSoftwareWindows 7/8/10, 32 or 64 bit
LIS SystemBi-direction, support HL7 protocol
InterfaceLAN port access
PrinterExternal optional, multiple reporting mode available
Working ConditionsPower SupplyAC220V±10%, 50/60Hz; 110V±10%, 60Hz, 300W
Temperature15~30℃
WaterConsumptionDeionized water: 5L/H
Humidity40%~85%
Size&WeightExternal Size (W*D*H)950*600*515mm
Net Weight75kg
Package Size(W*D*H)1130*735*1040mm
Gross Weight135kg